Battery charger
Model number
Mains voltage
Frequency
Permissible variation
Current consumption
Output voltage
Max. Output current
Sonopuls
Battery
Patient leakage current
Ditto, first fault condition
Safety class
Weight
Dimensions
Safety tests
Environment conditions for Transport and Storage
Environment temperature
Relative humidity
Atmospheric pressure
This equipment complies with all requirements of the Medical Device Directive
(93/42/EEC), Class IIb.
*II
: Safety class II (double insulated).
**BF : The equipment has a floating patient circuit. The Sonopuls 490 and Sonopuls 491
meets the requirements of IEC 601 -1, IEC 601-2-5 and IEC 601-2-1 0.
Copies of test reports available on application.
Technical modifications reserved.
For the ordering data of the Sonopuls, standard accessories and additional accessories we refer to the
Catalogue for Physiotherapy.
30
: ENA-1810
: 230 V EUR
230 V BSI (optional)
120 V UL/CSA (optional)
: 50/60 Hz
: ± 10%
: 17 VA
: 18 V
: 500 mA
: 12 V, 1.8Ah maintenance-free lead accumulator; dim.: 178.5 x 34 x
60.5 mm (l x w x h)
: typically 1 µA (requirement IEC ≤ 100 µA)
: typically 2 µA (requirement IEC ≤ 500 µA)
: II* type BF**, according to IEC 601-1
: 4.0 kg
: 290 x 280 x 105 mm
: TÜV Rheinland
: -10° till +50° C
: 10 till 100 %
: 500 till 1060 hPa
TECHNICAL DATA
Type BF
ORDERING DATA