TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the
socket.
PROBLEM
The device does not
work
The device does not
nebulize or
nebulizes insuffi-
ciently.
The apparatus is noi-
sier than usual
If after verifying the above mentioned conditions the device should not operate pro-
perly, contact your trusted retailer or the nearest authorized FLAEM service center.
CAUSE
The power cable has
not been correctly in-
serted into the power
socket
The medication has
not been inserted in
the nebulizer
The nebulizer has not
been properly fitted
The nebulizer nozzle is
clogged
The air pipe is not cor-
rectly connected to the
device
The air pipe is bent,
damaged or twisted
The air filter is dirty
The Filter is not cor-
rectly inserted in its
housing
8
SOLUTION
Correctly insert the power cable in
sockets.
Pour the right amount of medication
in the nebulizer
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on the
cover
Disassemble the nebulizer, remove
the nozzle and perform cleaning
operations. Failure to remove medi-
cation deposits from the nebulizer
affects its efficiency and operation.
Strictly comply with the instructions
contained in the "CLEANING, SANITI-
SATION, DISINFECTION" chapter
Ensure the device's accessories
are properly connected to the air
intake (see connection diagram on
the cover)
Unwind the pipe and ensure it is
not crushed or punctured. If neces-
sary replace it
Replace the filter
Fully and properly insert the Filter in
its housing