In general, when used in those concentrations recommended by the manufacturer, either phenolic type or quaternary
type disinfectants can be used. Iodophor type disinfectants are not recommended for use because staining may
result.
Suggested cleaners for the Power-PRO TL Cot Fastener System:
Quaternary Cleaners (active ingredient - ammonium chloride)
Phenolic Cleaners (active ingredient - o-phenylphenol)
Chlorinated Bleach Solution (5.25% - less than 1 part bleach to 100 parts water)
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer's guidelines
for proper disinfecting.
WARNING
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF
USED IMPROPERLY. If the products described above are used to clean Stryker patient care equipment, measures
must be taken to insure the equipment is wiped with a cloth soaked in clean water and thoroughly dried following

cleaning.

Failure to properly rinse and dry the equipment after cleaning may leave a corrosive residue on the surface of the
equipment, possibly causing premature corrosion of critical components.
Note: Failure to follow the above directions when using these types of cleaners may void this product's warranty.

REMOVAL OF IODINE COMPOUNDS

Use a solution of 1/2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area. Clean as soon
as possible after staining occurs. If stains are not immediately removed, allow solution to soak or stand on the surface.
Rinse surfaces which have been exposed to the solution in clear water before returning unit to service.
www.stryker.com
Cleaning
6385-009-005 REV B
Return To Table of Contents