5.4 Connection to the balance
Before connecting the second display or unplugging the second display unit to resp.
from the balance, make sure that the balance is unplugged from the mains.
This second display unit is exclusively suitable for the connection to KERN balances.
5.5 Initial Commissioning
In the menu of the connected balance in the menu item "F3 Con" the parameter "S
232 – P Cont" has to be set.
Set the baud rate at 9600 baud.
For details please see operating instructions of the connected balance.
Plug one end of the connecting cable into the balance and the other end into the
backside of the case of the second display unit.
Connect the second display unit with the delivered connection cable to the
balance
Power is supplied via the balance. The second display has no own power
supply unit.
The balance can now be controlled via the second display unit.
Press
6 Operation
6.1 Overview of display
6.2 Display
Turn on device using the
The appliance carries out a selftest and then changes to the zero display „0.0".
Afterwards the display of the connected balance will be imported. Now the second
display unit is ready for use.