Monthly service
Replacing the Cooling Water
Lubrication of Bearings for Tilting
Frame
Unitom-2
Instruction Manual
Replace the cooling water in the Recirculation Cooling Unit at
least once a month.
Press emergency stop.
Open the protection guard.
Remove the red protection caps from each of the two grease
nipples, located as indicated with a white circle in the illustration.
Lubricate in the two grease nipples with the grease gun and
move the spindle/cutting motor up and down by hand, until
grease comes out at the shaft. Use water-repelling grease, e.g.
Shell Albida Grease LX.
Clean superfluous grease away and mount the protection caps
at the grease nipples.
Release the emergency stop.
21