Safety; Electrical Safety - SonoSite Edge Stand Guide D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 49
a Use either a pre-moistened wipe or a soft cloth dampened with cleaner or disinfectant.
Approved cleaners and disinfectants for the stand
Cleaner/disinfectant
SaniCloth AF3 (gray top)
SaniCloth Plus (red top)
PI-Spray II
1
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a
minimum period of time.
2
Qualified for use as an intermediate-level disinfectant for mycobacteria. For a more complete list, refer to
the cleaners and disinfection document available at
b Remove all gel, debris, and bodily fluids from the stand.
c With a new wipe, clean all parts of the stand by wiping from the clean areas to the soiled areas. This
method helps to avoid cross-contamination.
d Observe the minimum wet contact time. Monitor the stand parts for wet appearance. Re-apply with a
new wipe if no longer wet.
3 Allow the stand to air dry in a clean, well-ventilated space.
For information about cleaning the peripherals, see the manufacturer's instructions.

Safety

Observe the following warnings and cautions before using accessories and peripherals with the ultrasound
system and transducers.

Electrical safety

The stand is not an electrical device and does not require ground bond testing. The stand is for facilitating the
mobility of peripherals and their connection to the ultrasound system.
12
Minimum wet contact time
2
3 minutes
3 minutes
10 minutes
1
www.sonosite.com
Safety

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

H-universal stand

Table des Matières