Cleaning and disinfecting
To avoid electrical shock, before cleaning always disconnect the system from the
WARNINGS
To avoid injury, always use protective eyewear and gloves when performing
To avoid infection:
Caution
Use only recommended cleaners or disinfectants on surfaces. Immersion-type
disinfectants are not tested for use on surfaces.
To clean and disinfect a peripheral, refer to its manufacturer's instructions. For the bar code scanner, see also
the Bar Code Scanner User Guide.
To clean and disinfect the stand or TTC
You can clean and disinfect the exterior surface of the stand or TTC using a recommended cleaner or
disinfectant. For the most recent list of approved products, see www.sonosite.com.
1 Turn off the ultrasound system and any accessory equipment. Disconnect the system and accessory
equipment from AC power.
2 Clean the surfaces of the stand to remove any debris or bodily fluids. Use the following procedure:
Cleaning and disinfecting
power supply and remove it from the stand.
cleaning and disinfecting procedures.
Ensure that the solution expiration date has not passed.
The level of disinfection required for a product is dictated by the type of tissue it
contacts during use. Ensure that the solution strength and duration of contact
are appropriate for the equipment. For information, see the disinfectant label
instructions and the recommendations of the Association for Professionals in
Infection Control and Epidemiology (APIC) and FDA.