Startup; Switching On; Köhler Illumination; Incident Light - Leica DM3 XL Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7. startup

7.1 Switching On

• Switch on the microscope at the on/off switch
(Fig. 2.7).
7.2 Köhler Illumination

7.2.1 Incident Light

The condenser module is pre-centered at the
factory. However, depending on how the stand
is transported and assembled, it may be neces-
sary to re-center the condenser in some cases.
Therefore, check the diaphragm centering.
The following procedure is provided for incident
light-bright field illumination.
• Select an objective with moderate magnifica-
tion (10x-20x).
• Switch on the BF (= brightfield) or Smith reflec-
tor.
• Place a specimen on the stage.
• Bring the specimen into focus and adjust the
light intensity.
Fig. 15 Köhler Illumination
1
Aperture diaphragm adjustment
2
Color coding
2
1
Adjusting the Aperture Diaphragm
• Use a flat specimen with good reflective prop-
erties, and a glass plate (specimen carrier) if
necessary.
• Remove one ocular (the right one, for exam-
ple). Alternatively, you can also use a focusing
telescope or Bertrand lens.
• Close the aperture diaphragm using the crank
(Fig. 15.1) until the edge of the diaphragm ap-
pears in the exit pupil of the objective (aperture
diaphragm plane).
• If the image does not appear in the middle of
the field of view, the aperture diaphragm must
be moved into the exit pupil using the centering
screws (Fig. 16.1).
• Open the aperture diaphragm until the color
ring of the objective and the color marking of the
aperture diaphragm (Fig. 15.2) match.
Fig. 16 Aperture diaphragm centering
1
Centering keys
2
Centering screws
1
2
7. Startup
25

Publicité

Table des Matières
loading

Table des Matières