Measuring protein contamination and monitoring cleaning
performance of surgical instruments
Items supplied
Storage
24
The instrument to be tested must be suitable for machine
reprocessing.
Please follow the manufacturer's advice in accordance
with the EN ISO 17664 standard.
– 48 individual bottles of reagent A
– 2 g bottle of reagent B with measuring spoon
– 5 ml of reagent C in a dropper bottle
– 100 ml bottle of rinsing solution = 5% SDS concentrate
– 1 pipette
– 1 cuvette
– Short instructions with colour chart for evaluation
– 2 test and cycle specific barcode strips for coding the
RQflex plus reflectometer
The test kit can be sealed and stored at room temperature
(15-30 °C) until the date given on the packaging.
Storage at higher temperatures (max. approx. 50 °C) for a
limited period of time (up to 3 months) will not impair the
performance of the test kit.
Low temperatures (below 10 °C) may cause the solution to
crystallise.
If this happens, heat the SDS solution to dissolve the
crystals. The solution must then be left to cool down to room
temperature again.
1 % ready-to-use SDS solution can be kept for 6 weeks at
room temperature or for 6 months in a refrigerator.