Knick Stratos Pro A2 COND Série Guide Rapide page 11

Masquer les pouces Voir aussi pour Stratos Pro A2 COND Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 25
Calibration with Calibration Solution
Display
Note:
• Be sure to use known calibration solutions and the respective
temperature-corrected conductivity values (see table on calibration
solution).
• Make sure that the temperature does not change during the
calibration procedure.
Action
Display of selected
process variable (here:
mS/cm). Now the de-
vice is in HOLD mode:
Reinstall the sensor
and check whether the
message is OK.
MEAS ends calibration,
REPEAT permits
repetition.
With MEAS selected:
Press enter to exit
calibration.
Remark
Display of conductiv-
ity and temperature,
Sensoface is active.
After end of calibra-
tion, the outputs
remain in HOLD
mode for a short
time.
After display of
GOOD BYE, the
device automatically
returns to measuring
mode.
11

Publicité

Table des Matières
loading

Table des Matières