WIKA 712.15.160 Mode D'emploi page 22

Masquer les pouces Voir aussi pour 712.15.160:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4. Commissioning, operation
Dismounting cable socket
The cable socket is prefitted and must be dismantled as described below.
EN
Screw
Cable cover cap
Cable cover
Socket insert
Cable socket baseplate
Sealings
Fully loosen the screws on the cable cover cap with a slotted screwdriver (0.6 x 3.5 mm)
and remove them.
Pull the cable cover, along with the socket insert, out from the cable socket baseplate
(joined permanently with the instrument).
Take the cable cover cap and push the socket insert out, downwards, completely through
the cable cap.
Pin assignment cable socket
4 ... 20 mA, 2-wire
Do not use this
terminal
U
B
Supply
voltage
+0V/GND
Evaluation
Terminals 3 and 4: For internal use only
(display)
Terminals 5 and 6: Reset zero point
With multiple scales or replaceable plug-in scales (option), the output signal of 4 ... 20 mA
corresponding to each scale can be stored in a microprocessor.
Differential pressure gauge models 712.15.160, 732.15.160, optional for hazardous areas
22
+/I+
Only use cable with a diameter of
7 ... 13 mm

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

732.15.160

Table des Matières