LAMPEGGIANTE
PRESS START
4S
TO CONTINUE
5
6
Heating
The device performs a heating phase lasting 5
minutes and a first decontamination of tools.
If temperature is not achieved, heating continues
also during the washing phase.
Degas (DEGAS)
The device activates intermittent ultrasounds for
5 minutes. This process reduces gases in the
fluid and improves effectiveness of the next
cleaning phase.
STANDBY
This new phase allows operating important
process as:
- Keeps the instruments in the washing liquid
(SOAKING)
- Add instruments
(ATTENTION: STEEL MAX 1 KG.)
TO EXIT THE STANDBY PHASE AND END
AUTOMATICALLY THE CYCLE IS
NECESSARY TO PRESS START.
Ultrasound cleaning
Cleaning phase by means of ultrasonic
cavitation.
SWEEP
Ultrasounds undergo frequency modulation thus
reducing the negative effects of stationary
waves and increasing distribution of ultrasonic
energy into the cleaning fluid.
Cleaning
is
more
improved)
Draining
The tank drains automatically
effective
(cavitation
ENGLISH
1
5 min
1
5 min
1 min
to
30 min
is
≈
8 min
13