9.1 Technical Data

Electrical data for 230VAC devices
Mains supply voltage
Mains frequency
Power consumption
Device fuse L/N
Electrical data for 115VAC devices
Mains supply voltage
Mains frequency
Power consumption
Device fuse L/N
Performance data
Suction category
Suction power max.
Vacuum meter
Vacuum regulator
Vacuum height max. at
Classification
Protection class
Degree of protection
Classification according MDD 93/42/EEC
Conformity for 230VAC devices
Concerning Medical Device Directive
MDD 93/42/EEC
Conformity for 115VAC devices
With respect to electric shock, fire and mechanical
hazards only in accordance with UL 2601-1 / 52YA.
With respect to electric shock, fire and mechanical
and other specified hazards only in accordance with
CAN/CSA C22.2 No.601.1, (CAN/CSA 601.2XX, if
applicable) Medical equipment certified for Canada
52YA E218915.
Ardo medical AG
I
N F O R M A T I O N
Master
Senator
Master
Senator
Master
Senator
2'000m a.s.l.
1'000m a.s.l.
500m a.s.l.
0m a.s.l.
0123
230VAC
50Hz
140W / 0.6A
115W / 0.45A
T1.25AH / AC250V
115VAC
60Hz
140W / 1.2A
115W / 1.0A
T1.60AH / AC250V
high vacuum / high flow
50l/min ±15%
30l/min ±15%
0 to -100kPa / 0 to -700mmHg / ±2.5%
mechanically, continuously 0 to maximum
-72kPa / -540mmHg / ±5%
-84kPa / -630mmHg / ±5%
-89kPa / -668mmHg / ±5%
-95kPa / -713mmHg / ±5%
Class I with protective earth ground
BF
IIa
EN 60601-1 / EN60601-1-2 / EN60601-1-6 /
EN ISO 10079-1 / EN ISO 14971
E218915
UL2601-1
52YA
19