9. Troubleshooting

Error
No display.
Cannot change
cleaning time.
Cannot save
cleaning time.
Insufficient
cleaning
performance.
Cleaning needles
have corroded.
Cause
Appliance not connected to
mains.
Appliance not turned on.
No power from the socket.
Faulty fuse.
Appliance still running.
Cleaning time was entered
during an interrupted
cleaning process.
Cleaning time too short.
Active ingredients of the
activefluid cleaning liquid
spent.
Insufficient number of‚
cleaning needles.
Cleaning needles spent.
Cleaning needles have been
insufficiently rinsed.
Cleaning needles have been
stored incorrectly.
Cleaning needles in use for
longer period of time.
Solution
Check correct fit of power cable in the
appliance socket and wall socket.
Turn on appliance at the mains switch
(L, Fig. 3).
Check power socket, for example, with
another appliance.
Replace the fuse (see section 6.3.). If the
fuse blows repeatedly, have the appliance
inspected by a service technician.
Interrupt the cleaning process with the
Start/ Pause button, and then adjust
cleaning time.
Stop the cleaning process with the Stop
button, and then enter new cleaning time.
See chapt. 5.5.
Prepare fresh activefluid cleaning liquid
after each application.
Fill up the cleaning needles
(see section 5.7.2.).
Replace the cleaning needles
(see section 6.1.).
Rinsing and storage see section 5.7.1.
Removing the corrosion:
Clean the needles with 100 ml activefluid
cleaning liquid, without contents, for
5 min at 2000 min
.
-1
Dispose of activfluid cleaning liquid.
The needles are so-called consumable
goods. If it is not possible to regenerate
the needles or the needles' performance
deteriorates, see chapter 6.1.
activeblue
EN
en
29