Concentration main alarm A2
Danger to life! Leave the area immediately.
A main alarm is self-latching and cannot be
acknowledged (see Dräger X-am 5000 Instructions
for Use).
The alarm is indicated by an
intermittent alarm message:
The triggering device at A2:
– A double tone sounds and the LED ring flashes red
twice (main alarm).
A receiving device at A2:
– A double tone sounds and the LED ring flashes red/
green twice (secondary alarm).
For O
: A1 = lack of oxygen
2
A2 = excess oxygen
Before the area may be entered again, a clearance
measurement must be performed!
Only after the concentration has dropped below the
alarm threshold A2:
Press
key on the Dräger X-am 5000 of the
OK
triggering Dräger X-zone 5000.
– The alarm messages will be switched off.
Battery pre-alarm
The alarm is indicated by an
intermittent alarm message:
Acknowledging the pre-alarm:
– The Battery LED flashes red.
Press
key on the Dräger X-am 5000 of the
OK
Dräger X-zone 5000, only the acoustic alarm is
switched off.
– After the first battery pre-alarm the battery will last for
another approx. 15 minutes.
Battery main alarm
The alarm is indicated by an
intermittent alarm message:
Only the acoustic signal of the main alarm can be
acknowledged.
– The Battery LED flashes red.
– The device automatically switches off after
10 seconds.
The visual and the audible alarm will be activated for
a short time before the device switches off.
DANGER
WARNING
Device alarm
NOTICE
Indicates a malfunction of the Dräger X-zone 5000 or
the Dräger X-am 5000.
The alarm is indicated by an
intermittent alarm message:
– The device or one or more sensor channels are not
ready for operation.
– Remedies, refer to "Faults, cause, remedy" on
page 40
If necessary, commission Dräger to eliminate the
error.
Acknowledge the device alarm
Press the
key on the Dräger X-am 5000 or on
OK
the Dräger X-zone 5000.