Communication Fieldbus / appareil de mesure
Paramètres de bus / Blocs Analog Input (AI)
Parameter
Index
1
ST_REV
2
TAG_DESC
3
STRATEGY
4
ALERT_KEY
5
MODE_BLK
6
BLOCK_ERR
7
PV
8
OUT
9
SIMULATE
10
XD_SCALE
11
OUT_SCALE
12
GRANT_DENY
IO_OPTS
13
14
STATUS_OPTS
15
CHANNEL
16
L_TYPE
17
LOW_CUT
18
PV_TIME
19
FIELD_VAL
UPDATE_EVT
20
94
Description
Static Revision
TAG Description
Strategy
Alert Key
Target
Actual
Permitted
Normal
Block Error
Process Value
Status
Measured Value
Status
Simulate Status
Simulate Value
Transducer Status
Transducer Value
Simulate Enable/ Disable
High Range
Low Range
Units Index
Decimal Point
High Range
Low Range
Units Index
Decimal Point
Grant
Deny
IO Block Options
Status Options
Channel
Linearization Type
Low Cut Off
Filter Time
Percent Value
Status
Unacknowledged
Update State
Time Stamp
Static Revision
Relative Index
Transmetteur Cond Ind 7100e FF
R/W
Default
0
R
R/W
0
R/W
0
R/W
OOS
R/W
-
OOS, Auto
Auto
R
R
R
R
R
R/W
R/W
R
R
R/W
100
R/W
0
R/W
0
R/W
0
R/W
100
R/W
0
R/W
0
R/W
0
R/W
0
R/W
0
R/W
0
R/W
1
R/W
0
R/W
0
R/W
0
R/W
R
R
0
R/W
0
R
0
R
0
R
0
R