FreeStyle Comfort
7
Low
Concentration
Priority #
Alarm
Description
8
System
Malfunction
Flow Rate
(OCD Valve
Failure,
Obstruction of
Gas Pathway)
9
Low Battery
Warning
10
Breathing
Rate
Exceeded
11
System
Malfunction
Fan Failure
12
System
Malfunction
Low Pressure
13
System
Malfunction
Low Temp
Info
Start up
20 - ENG
PN MN250-C4 G | User Manual
Display
1 beep
If problem persists, contact provider.
every
30 sec,
for up to
5 min
Audible
More Details
Signal
1 beep
If problem persists, contact provider.
every
30 sec
1 beep, no
Charge battery.
repeat
1 beep
Breathing Rate exceeds 40 bpm.
every
If problem persists, contact provider.
60 sec
none
If problem persists, contact provider.
none
If problem persists, contact provider.
none
Allow unit to warm up to room temperature.
If problem persists, contact provider.
1 info
Screen displays at power-on and you should
beep
hear one beep to confirm buzzer is functioning
Start up screen display current firmware
revision in bottom right corner.