Maintenance and Troubleshooting
8.7
Service Program for moisture removal in the vacuum pump
Important !
Please activate the Service Program once a week.
To remove moisture from the pump, the Service Program can be used. In approx. 30
minutes it will warm up the pump and remove the condensed water from the oil. The
following procedure should be followed.
Activation of Service Program:
Press and hold the arrow buttons simultaneously.
Switch on the machine using the master switch.
Close the chamber cover
 the Service Program starts (duration approx. 30 minutes).
 Display shows SE.
If desired: Abort the Service Program by pressing