DeVilbiss Healthcare Drive ENVOY Mode D'emploi page 26

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

EN
Error code description S-Drive
If the red LED in the battery display is blinking, the error can be determined using the blink code:
Number of
Error
blinks
1
Low battery voltage
2
Motor not connected
3
Motor wiring problem
4
Freewheel activated
5
not allocated
6
Lock active
7
Error of the rotational
speed potentiometer
8
Control unit defective
9
Parking brake
10
Battery voltage is too high Driving is not
26
Drive Medical GmbH & Co. KG | As per: 10.08.2020 | Errors and alterations excepted
Effect on the
Comments
scooter
Driving is not
The charge status of the batteries is low. Charge the bat-
possible
or
teries immediately. If there is no charging, check the bat-
slow
tery cables to ensure that they are positioned securely
and connected properly.
Driving is not
The contact between the motor and the control unit is
possible
poor; check all plug connections from and to the motor.
Driving is not
The motor or its cable is defective. Check exposed
possible
cables or short-circuited electrical circuits. Contact your
customer service.
Driving is not
Check whether the gearshift lever for the parking brake
possible
is in the locked position. The brake coil or the cables of
the parking brake are defective. Check the parking bra-
ke and the cables for short-circuited electrical circuits or
exposed cables.
Driving is not
The charging unit is still connected. Please disconnect
possible
the charging unit from the vehicle.
Driving is not
The throttle lever or its cable is defective. Check ex-
possible
posed cables or short-circuited electrical circuits. The
throttle lever could be incorrectly adjusted. Contact
your customer service.
Driving is not
Check the plug connections on the control unit; possibly
possible
replace the control unit.
Driving is not
The brake coil or the cables of the parking brake are
possible
defective. Check the parking brake and the cables for
short-circuited electrical circuits or exposed cables.
The charge status of the batteries is too high. Pull out
possible
the charging unit connection if it is connected or rotate
the "Charge/Run" switch into the "Run" position. Scoo-
ters which are operated with RHINO charge batteries
while the scooter rolls down an inclined surface or dece-
lerates. This error can be caused by excessive charging
in this manner. Switch the scooter off and then back on.
Instructions for use ENVOY

Publicité

Table des Matières

Dépannage

loading

Table des Matières