System Set-Up; Connecting The Accessories; Inserting The Rechargeable Battery; Intraoperative Battery Change - B.Braun Aesculap Acculan 4 GA344 Mode D'emploi/Description Technique

Petite perceuse
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Aesculap
Small drill GA344
Risk of injury and material damage due to inappro-
priate use of tools!
WARNING
Always follow the safety advice and information
given in the instructions for use.
When coupling/uncoupling, handle tools with
cutting edges with care.
Damage to the product if dropped!
Use the products only if they are in perfect
technical condition, see Function check.
WARNING
Risk of burns to skin and tissue through blunt tools
or if product has not been maintained properly!
Use tools only if they are in perfect condition.
WARNING
Replace blunt tools.
Maintain the product properly, see Mainte-
nance.
6.1

System set-up

6.1.1

Connecting the accessories

Risk of injury due to unapproved configuration
using additional components!
Ensure that the classification of all components
DANGER
used is consistent with the classification of the
product (such as type BF or type CF).
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
All
configurations
must
IEC/EN/DIN 60601-1. The person connecting the devices with each other
is responsible for the configuration and must ensure that the fundamental
standard IEC/EN/DIN 60601-1 or relevant national standards.
Follow the instructions for use of individual accessories.
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
6.1.2

Inserting the rechargeable battery

Turn product 1 with the battery shaft 4 upwards and attach the sterile
funnel 6 (sterile), see Fig. A.
Insert battery 7 (non-sterile) into the battery shaft 4 by a second (non-
sterile) person, see Fig. A.
6
Acculan 4
®
fulfill
the
fundamental
Note
After the battery has been inserted, multiple signals will sound, which indi-
cate that the product is ready to use.
After insertion of the battery, have the sterile funnel 6 (non-sterile)
removed by a second (non-sterile) person.
Attach the cap 8 (sterile) in such a manner that it clicks into place with
both release catches 9.
Note
The sterility of the product is only guaranteed with a correctly placed cap.
6.1.3

Intraoperative battery change

WARNING
The battery removal device is used for intraoperative battery changes
ensuring that sterile conditions are safely maintained.
Turn the product 1 with the battery shaft 4 upwards.
Press both safety catches 9 on the cap 8 at the same time and remove
the cap 8.
Attach the sterile battery removal device 10, see Fig. B.
Shake the product 1 with the attached battery removal device 10 with
battery shaft 4 slightly downwards.
The battery 7 will slide easily into the battery removal device 10.
Give the battery removal device 10 including dead battery 7 to the
non-sterile person.
Insert a charged battery 7, see Inserting the rechargeable battery.
6.1.4

Removing the rechargeable battery

standard
WARNING
CAUTION
CAUTION
Risk of injury from hot battery!
Battery may be hot after use in the machine.
Remove the battery with the battery removal
device and let it cool.
Risk of injury from hot battery!
Battery may be hot after use in the machine.
Let the battery cool in the machine and then
remove.
– or –
Remove the battery with the battery removal
device and let it cool.
Damage to the battery caused by knocking it
against hard objects!
Only remove battery by tapping the product on
the palm of the hand.
Damage to, or destruction of the batteries due to
processing!
Do not sterilize batteries.

Publicité

Table des Matières
loading

Table des Matières