Supplementary Instructions on Processing marCore
MIS Instruments
These additional instructions are intended only for marCore
numbers ranging from REF 24-599-00-04 to REF 24-599-99-04.
The instructions provided in the foregoing sections of this document remain fully valid for
®
marCore
MIS instruments as well. They are merely supplemented as follows.
7.1

Product Overview

Fig. 7-1:
Product overview
Fig. 7-2:
Instrument – example
1 Working end
2 Shank
3 Handle
The handle (3) and the nut (4) are made of titanium and have been color-coded. Gentle
cleaning of these parts requires the use of neutral-pH or mildly alkaline detergents. The
shank (2) and the working ends (1) of these instruments are stainless steel with a protective
metal coating that increases the service life of the instruments a great deal as long as they are
used as intended. The cleaning connector (5) of the instruments is also stainless steel.