Input
Resistive sensor
Measuring ranges
Connection
Max. line resistance
Supply current
Output Data
Output
Maximum load
Residual ripple
Input unconnected
or measuring range
exceeded
Transmission Behavior
Transmission error
Temperature influence
Time response
Pt100 acc. to DIN 60751
P44100D1-0004
P44100D1-0005
P44100D1-0006
P44210D3-0007
P44210D3-0008
P44210D3-0009
2-, 3- or 4-wire
Note: With 3-wire connection, the sensor
cable resistance is not completely compen-
sated for.
100 ohms
Approx. 1 mA
4 ... 20 mA (linear up to 21 mA)
550 ohms
< 10 mV
rms
> 21 mA (max. 38 mA)
± 1 K (typically ± 0.5 K)
at 23 °C ambient temperature
< 150 ppm/K of full scale
(average TC within allowable operating temp range,
reference temp 23 °C)
T
time max. 100 ms
90
Specifications
0 ... 150 °C
0 ... 200 °C
0 ... 300 °C
0 ... 150 °C
0 ... 200 °C
0 ... 300 °C
33