KLS Martin group marVac smart Vac Mode D'emploi page 52

Détecteur de courant de réseau externe idetect
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

If the fan of the marVac
Repeat the calibration process. See section 5.3, "Calibration," page 47. If this does not produce the
desired improvement: Increase the L-parameter step by step.
If the line current detector does not activate the fan, even though the connected electrosurgery or laser
surgery device is activated: Reduce the L parameter step by step.
Electro-coagulation currents may require tissue contact to activate the marVac
Recommended values for the switch-off threshold
Gebrüder Martin ME 402
Gebrüder Martin ME MB1
Gebrüder Martin ME 411
Gebrüder Martin MCO 25 plus
Gebrüder Martin MCO 50 plus
general
A small L parameter (<15%) can be selected for electrosurgery or laser surgery devices with a high power
consumption; a larger L parameter (>20%) can be selected for devices with a low power consumption.
Large L parameters reduce the probability of unwanted fan activation.
52
®
runs automatically while the connected surgical device is in operational status:
15% – 20%
20% – 30%
20% – 25%
15%
15%
15%–25 %
marVac® / maXium® smart Vac iDetect external line current detector
Instructions for Use
®
fan.
Revision 04

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Maxium smart vac

Table des Matières