13.1. Ultrasound

Frequencies
Display
Contact control threshold
Timer
Ultrasound, continuous
Pulse frequency / duty cycle
Intensity
Ultrasound, pulsed
Pulse frequency / duty cycle
Intensity

13.2. Treatment head

1 MHz, large
* SATP = Spatial Average Temporal Peak (average pulse power)
** ERA
= Effective Radiating Area, this is the effective radiating area of the treatment head
***BNR
= Beam Nonuniformity Ratio, indicates the ratio between the peaks and the average value of the intensity

14.1. Mains adapter

Model number
Mains voltage
Frequency
Max, Power consumption

14.2. Sonopuls

Patient leakage current
idem single fault condition
Safety class
Dimensions
Weight
Certificates
CE marking
Class
Safety standards
26
: Geometric surface 6,2 cm², ERA** 5,0 cm ², BNR*** max. 6 type collimating,
side panel radiation max. 10 mW/cm²
in the sound beam. A low BNR excludes high, unwanted energy concentrations.
: 1 MHz
: intensity in W/cm² and power output in W (SATP*)
: 65 %
: 0 - 30 minutes, coupled to contact control
: 100 Hz / 100 %
: 0 - 2 W/cm²
: 100 Hz / 5, 10, 20, 50, 80 %
: 0 - 3 W/cm², duty cycle 5, 10, 20, 50 %
: 0 - 2 W/cm², duty cycle 80 %
:
ENB-1530
:
100 - 240 Volt
:
50/60 Hz
:
40 VA
:
typically 1 µA (requirement IEC < 100 µA)
:
typically 2 µA (requirement IEC < 500 µA)
:
II* type BF**, according to IEC 60601-1
:
220 x 200 x 195 mm
:
1.7 kg
:
CB / GOST
:
***
:
IIb ***
:
IEC 60601-1 and IEC 60601-2-5.