8. Dismounting, return and disposal / 9. Specifications

8.3 Disposal

Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specific waste disposal regulations.
Dispose of the hydraulic fluids as described in the material safety data sheet.
The current material safety data sheet can be found at www.wika.de on the
product page of the respective product.
9. Specifications
Specifications
Pressure range
Pressure transmission
medium
Fluid reservoir
Pressure connections
For reference instrument
For test item
Fine adjustment
Material
Sealings
Dimensions (L x W x H)
Weight
1) Other pressure transmission media on request.
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 91.07 and the order documentation.
WIKA operating instructions, models CPP700-H and CPP1000-H
Model CPP700-H
0 ... 700 bar
(0 ... 10,000 psi)
Hydraulic fluid based on mineral oil or single distilled water
(water-based hydraulic fluids are not suitable)
200 cm³
G ½ female thread, freely rotating.
The connection can be disconnected using an Allen key. Then a
G ⅜ female thread is available.
G ¼ female at test connection
hose, length 1 m (3.28 ft),
®
Minimess
1620 system
Fine adjustment valve / volume adjustment
anodised aluminium, brass, stainless steel, ABS
FKM and NBR (standard)
280 x 170 x 120 mm
(11.0 x 6.7 x 4.7 in)
1.9 kg
(4.2 lbs)
Model CPP1000-H
0 ... 1,000 bar
(0 ... 14,500 psi)
G ¼ female, freely rotating at
test connection hose, length
1 m (3.28 ft)
EN
1)
23