Troubleshooting

This troubleshooting documentation details isolation to either the
probe or the TEKPROBE oscilloscope.
To troubleshoot the P6245, the probe must be attached to an
operating TEKPROBE interface oscilloscope with the top cover of
the compensation box removed. Refer to page 4- - 14 for instructions
on opening the compensation box.
The host instrument should be set to the following settings:
H 50 Ω termination at instrument.
H DC coupling
CAUTION.
box. Operating voltages will be present. Be careful not to touch
or ground energized components.
Work at a Level 1 ESD workstation when performing troubleshoot-
ing. The compensation box contains static sensitive devices.
P6245 Instruction Manual
Observe caution when probing in the compensation
4- 19