DEVILBISS GUIDANCE AND MANUFACTURER'S DECLARATIoN
wARNING
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying
documents.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining
the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the
Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing
other Equipment and Systems or non-medical electrical equipment.
Guidance and Manufacturer's Declaration – Emissions All Equipment and Systems
This device is intended for use in the electromagnetic environment specified below. The customer or user of this device
should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
Class B
RF Emissions
Radiated and Conducted
CISPR 11
Emissions
Harmonics
Class A
IEC 61000-3-2
Flicker
Complies
IEC 61000-3-3
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC 61000-4-2
Electrical Fast
Transient/burst
±2kV on AC Mains
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Voltage dips, short
>95% Dip for 0.5 Cycle
interruptions and
60% Dip for 5 Cycles
voltage variations
30% Dip for 25 Cycles
on power supply
input lines
>95% Dip for 5 Seconds
IEC 61000-4-11
A-7310
Electromagnetic Enforcement – Guidance
This device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
This device is suitable for use in all establishments including domestic,
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Compliance Level
±6kV contact
±8kV air
±2kV on AC Mains
±1kV Differential
±2kV Common
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of this device requires continued operation
during power mains interruptions, it is
recommended that the device be powered from
an uninterruptible power supply or battery.
EN - 9