Guidance and manufacturer's declaration –
The Aeroneb Solo nebulizer system is intended for use in the
electromagnetic environment specified below. The customer or the
user of the Aeroneb Solo nebulizer system should assure that it is
used in such an environment.
immunity
iec/en
test
60601 test
level
Conducted
3 Vrms
RF IEC
150 kHz to
61000-4-6
80 MHz
Radiated
3 Vrms
RF IEC
80 MHz to
61000-4-3
2.5 GHz
Aeroneb
Solo System Instruction Manual
®
electromagnetic immunity
compliance
level
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Aeroneb Solo
nebulizer system, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
recommended separation
distance
d = [1.17] √P
[3]V
d = [1.17] √P... 80MHz to 800MHz
d = [2.33] √P... 800MHz to 2.5GHz
where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
[3]V
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, ª should be less than
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
electromagnetic
environment – guidance
b