Specifications - Stryker TP600 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Classification
Size (approx.)
Weight
Reservoir capacity
Flow rate
Ambient operating temperature
Environmental conditions for transport
and storage
Temperature setpoints
Average temperature accuracy
Maximum Contact Surface
Temperature
High Limit Safety Temperature
Power cord
Current leakage
Ground resistance
Electrical requirements
Voltage (VAC)
Frequency (Hz)
Current (amps)
Certifications

Specifications

Class I equipment with Type BF applied part suitable for continuous operation. Not
classified for protection against ingress of liquid. Not classified for use in the presence
of flammable anesthetics.
11.5" x 8" x 8"
29.2cm x 20.3cm x 20.3cm
6.5 lb (2.9 kg) when empty
9 lb (4.0 kg) with unit filled with water to heating level
93 oz (2.75 l) maximum
7 gph (26.5 lph) minimum with pad attached
60°F to 90°F (15.6°C to 32.2°C )
-20°F to 120°F (-28°C to 48°C )
At uncontrolled RH
TP650 Classic
107°F (42°C) High heat
100°F (38°C) - Low heat
50°F (10°C) - Cooling
TP700 Professional
107°F (42°C)
100°F (38°C)
95°F (35°C)
50°F (10°C)
±2°F at 107°F (±1°C at 42°C)
107°F (42°C)
110°F to 120°F
(43.3°C to 49°C)
International (harmonized) 3-wire cordset
100 microamperes maximum
0.5 ohm max
120±10%
60
3.1 amperes
MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE,
AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1, AND CAN/CSA
C22.2 NO 601.1, ASTM F 2196-2002
IEC 60601-1-2
101515 REV C
1-21