5.6 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each
balance must be coordinated - in compliance with the underlying physical weighing principle
- to the existing acceleration due to gravity at its place of location (only if the balance has
not already been adjusted to the location in the factory). This adjustment process must be
carried out during the initial start-up, after change in location and variation of surrounding
temperature. To receive accurate measuring values it is also recommended to adjust the
balance periodically in weighing operation.
5.7 Adjustment
With an adjustment weight, the weighing accuracy can be checked and re-adjusted at any
time.
Attention: In the verified balances the adjustment is not possible.
Procedure when adjusting:
Observe stable environmental conditions. A short warming up time of ca. 15 minutes is
recommended for stabilization.