Certificate Of Compliance - Dräger PIR 7000 Notice D'utilisation

Masquer les pouces Voir aussi pour PIR 7000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36

Certificate of Compliance

1839123 (097594_0_000)
Certificate:
Project:
70055568
Issued to:
Draeger Safety AG & Co. KGaA
Revalstrasse 1
Luebeck, 23560
GERMANY
Attention:
Thomas Treptow
The products listed below are eligible to bear the CSA Mark shown
with adjacent indicators 'C' and 'US' for Canada and US or with adjacent
indicator 'US' for US only or without either indicator for Canada only.
PRODUCTS
CLASS - C482801 - SIGNAL APPLIANCES --Combustible Gas Detection Instruments-For Hazardous Locations
CLASS - C482881 - SIGNAL APPLIANCES-Combustible Gas Detection Instruments-For Hazardous Location-
Certified to U.S. Standards
For details related to rating, size, configuration, etc. reference should be made to the CSA Certification Record or
the descriptive report.
Class I, Div. 1, Groups B, C and D; Class II, Div. 1, Groups E, F and G;
Class I, Zone 1, Group IIB + H2
Type 4X
Model IDS0101, stationary, input rated 9 to 30 V dc, 9 W, output rated 4-20 mA. May be used with splash guard,
flow cell and bump test adapter. Temperature code of T6 for -40°C to + 40°C and T4 for -40°C to + 80°C.
Methane, propane and ethylene versions performance tested.
Model IDS0102, stationary, input rated 9 to 30 V dc, 9 W, output rated 4-20 mA. May be used with splash guard,
flow cell and bump test adapter. Temperature code of T6 for -40°C to + 40°C and T4 for -40°C to + 80°C.
Propane and methane versions performance tested.
CLASS 4828 02 - SIGNAL APPLIANCES - Toxic Gas Detection Instruments - For Hazardous Locations
DQD 507 Rev. 2012-05-22
Page 1
474
Master Contract: 160220
Date Issued: 2016-09-19
Zahra Amini
Issued by:
Zahra Amini
1839123
Certificate:
70055568
Project:
CLASS 4828 82 - SIGNAL APPLIANCES - Toxic Gas Detection Instruments - For Hazardous Locations -
CERTIFIED TO U.S. REQUIREMENTS
Class I, Div. 1, Groups B, C and D; Class II, Div. 1, Groups E, F and G;
Class I, Zone 1, Group IIB + H2
Type 4X
Model IDS0105, stationary, input rated 9 to 30 V dc, 9 W, output rated 4-20 mA. May be used with splash guard,
flow cell and bump test adapter. Temperature code of T6 for -40°C to + 40°C and T4 for -40°C to + 80°C.
WARNING: "The CSA certification does not cover the gas detection performance in Class II environments. The
sensor may become clogged and may not detect gas. In this event, the detector will signal a beam block fault."
APPLICABLE REQUIREMENTS
CAN/CSA-C22.2 No. 0-M91
General Requirements – Canadian Electrical Code, Part II
C22.2 No. 25-1966
Enclosures for Use in Class II, Groups E, F and G Hazardous Locations
CSA Std C22.2 No.30-M1986 Explosion-Proof Enclosures for Use in Class I Hazardous Locations
CSA Std C22.2 No.152-M1984 Combustible Gas Detection Instruments
UL50, Eleventh Edition,
Enclosures for Electrical Equipment
UL 1203 (4th Ed.)
Explosion-Proof and Dust-Ignition-Proof Electrical Equipment for Use in
Hazardous (Classified) Locations
IEC E60079-0:04
Electrical apparatus for explosive gas atmospheres - Part 0: General
Requirements
IEC E60079-1:05
Electrical apparatus for explosive gas atmospheres - Part 1: Flameproof
Enclosures "d"
CENELEC EN 60079-0:06
Electrical apparatus for explosive gas atmospheres - Part 0: General
Requirements
CENELEC EN60079-1:04
Electrical apparatus for explosive gas atmospheres - Part 1: Flameproof
Enclosures "d"
ISA-12-13.01-2003
Performance Requirements for Combustible Gas Detectors
CAN/CSA-C22.2 No.
Safety Requirements for Electrical Equipment for Measurement, Control,
61010-1-12
and Laboratory Use, Part 1: General Requirements
ANSI/ISA-61010-1
Safety Requirements for Electrical Equipment for Measurement, Control and
3rd Edition
Laboratory Use - Part 1: General Requirements - Third Edition
DQD 507 Rev. 2012-05-22
Page 2
Master Contract: 160220
Date Issued: 2016-09-19

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Pir 7200

Table des Matières