For validation purposes please make sure that you follow the
loading instructions given on the template.
Before loading the machine and before starting a
programme
– Are the spray arms able to rotate freely?
– Check spray arm jets for blockages.
– Check that the mobile unit is correctly docked to the water path in
the chamber.
Programme selection
When selecting a programme manually on the machine, the relevant
mobile unit must be selected first. Pictograms representing the mo-
bile units are shown on the display.
The pictogram can be found in the bottom right on the front of the
mobile unit, next to the RFID tag.
For the A 506 mobile unit, the pictogram below is displayed.
Once you have selected the pictogram, the list of assigned repro-
cessing programmes is displayed.
Mobile unit
A 506
en - Areas of application
Pictogram
31