E ngli sh
*THESE CHECKS SHOULD INCLUDE:
16. QUICK RELEASE PIN: Ensure the pin is securely fitted before lifting, by
pressing the barbed end.
17. LIFTING STRAP: Your lift strap should be checked at service intervals for
any sign of wear (ideally every six months). In order to view the strap it is
recommended to remove the cap at the top of the mast, cycle the lift in both
directions whilst observing the condition of the strap. If you see any wear
(fraying, creasing or similar) it is recommended an authorised, Joerns
Healthcare Dealer or qualified, competent service engineer replaces the
strap IMMEDIATELY. For details regarding strap maintenance and
replacement please request our detailed service guide. (See below for details)
Check
Lift Strap
Oxford
Stature
ü
ü
ü
ü
Check
Lift Strap
294000.10017 Rev. C
®