Screwing In Nozzles; Securing Injector Nozzles; Test Point For Measuring Water Pressure - Miele professional A 105/1 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 78
Securing injector
nozzles

Screwing in nozzles

Screw the injector nozzles into the required positions.
Use the 9 mm spanner (WAF 9) to tighten the injector nozzles.

Test point for measuring water pressure

The connection at the top of the right-hand injector manifold is used
as the test point for measuring water pressure. This connection point
can be used for testing pressure within the framework of performance
and validation testing in accordance with EN ISO 15883.
To measure the water pressure, replace the existing fixture with a
Luer Lock adapter, e.g. E 447.