Spécifications d'étalonnage ISO 8655
SpécificationsThermo Scientific Finnpipette F2 selon la norme ISO 8655.
Finnpipette F2 monocanal à volume variable
Gamme
Volumes
0,2-2 µl
0,5-5 µl
1-10 µl
2-20 µl
5-50 µl
10-100 µl
20-200 µl
100-1000 µl
0,5-5 ml
1-10 ml
Finnpipette F2 monocanal à volume fixe
Fixed Volume
µl
1
5
10
20
25
50
100
200
250
500
1000
2000
3000
5000
10000
Finnpipette F2 multicanaux à volume variable
Gamme
Conduit
1-10 µl
8, 12, 16
5-50 µl
8, 12, 16
10-100 µl
8, 12
30-300 µl
8, 12
Erreur de justesse
µl
µl
2
±0.080
0.2
±0.080
5
±0.125
0.5
±0.125
10
±0.120
1
±0.120
20
±0.20
2
±0.20
50
±0.50
5
±0.50
100
±0.80
10
±0.80
200
±1.60
20
±1.60
1000
±8.0
100
±8.0
5000
±40.0
500
±40.0
10000
±60.0
1000
±60.0
Erreur de justesse
µl
%
±0.050
±5.00
±0.125
±2.50
±0.120
±1.20
±0.20
±1.00
±0.50
±2.00
±0.50
±1.00
±0.80
±0.80
±1.60
±0.80
±4.00
±1.60
±4.00
±0.80
±8.00
±0.80
±16.0
±0.80
±40.0
±1.33
±40.0
±0.80
±60.0
±0.60
Volumes
Erreur de justesse
µl
µl
10
±0.24
1
±0.24
50
±1.0
5
±1.0
100
±0.80
10
±0.80
300
±8.0
30
±8.0
32
Erreur de répétabilité
%
s.d. µl
±4
0.040
±40
0.040
±2.5
0.075
±25
0.075
±1.2
0.080
±12
0.080
±1.0
0.10
±10.0
0.10
±1.0
0.20
±10
0.20
±0.8
0.30
±8.0
0.30
±0.8
0.60
±8.0
0.60
±0.8
3.0
±8.0
3.0
±0.8
15.0
±8.0
15.0
±0.6
30.0
±6.0
30.0
Erreur de répétabilité
s.d.µl
0.050
0.075
0.080
0.10
0.20
0.20
0.30
0.60
1.50
1.50
3.00
6.00
15.00
15.00
30.00
Erreur de répétabilité
%
s.d.µl
±2.4
0.16
±24
0.16
±2.0
0.4
±20
0.4
±0.8
0.30
±8.0
0.30
±2.7
3.0
±26.7
3.0
cv%
2.0
20.0
1.5
15
0.8
8.0
0.5
5.0
0.4
4.0
0.3
3.0
0.3
3.0
0.3
3.0
0.3
3.0
0.3
3.0
cv%
5.00
1.50
0.80
0.50
0.80
0.40
0.30
0.30
0.60
0.30
0.30
0.30
0.50
0.30
0.30
cv%
1.6
16
0.8
8.0
0.3
3.0
1.0
10.0