3. Specifications
Response time
T05 < 5 s (per DIN EN 60751)
T09 < 10 s (per DIN EN 60751)
EN
Maximum operating pressure
150 bar (2,175 psi)
When using a compression fitting:
max. 50 bar at 120 °C (max. 725 psi at 248 °F)
3.6 Accuracy data
Analogue signal
≤ ±0.5 % of span + temperature sensor error
Switching output
≤ ±0.8 % of span + temperature sensor error
Display
≤ ±(0.8 % of span + temperature sensor error) ±1 digit
Temperature sensor
For °C: ±(0.15 K + 0.002 | t |) per EN 60751
For °F: ±[1.8*(0.15 + 0.002 (t - 32) / 1.8)]
|T| is the value of the temperature without consideration of the sign.
The actually achievable accuracy is significantly determined by the mounting situation (immersion
depth, sensor length, operating conditions). This is especially the case for large temperature gradients
between the environment and the medium.
12
WIKA operating instructions temperature switch, model TSD-30