Product Description; Intended Purpose; Additional Regulations For Pressure Regulators With Pressure Outlet Apply; Additional Regulations For Pressure Regulators With Flow Outlet Apply - Emerson Tescom WegaMed Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

2

PRODUCT DESCRIPTION

2.1

Intended purpose

en
The pressure regulator is used for a variety of medical equipment in the field of medicine
and therapy to regulate the pressure at which medical gases and/or gas mixtures are
administered.
For connection by a trained user to a compressed gas cylinder.
Only for the medical gas or test gas mixture specified on the pressure regulator if
the pressure regulator does not have any gas type-specific identifying marking (see
Section 5.1).
Medical gases and test gas mixtures are a medicinal product and may only be
administered under the supervision of qualified medical staff who are familiar
with the precautions for the respective application. The safety data sheets and
information provided by the gas manufacturers must be observed.
For supplying suitable medical equipment in accordance with the technical
specifications indicated on the type plate and in the instructions for use.
For stationary use in hospitals and clinics.
• Not suitable in use with hyperbaric pressure chambers!
• Not suitable as a drive for surgical equipment!
• Not suitable for connection to medical supply units and its parts!
• Not suitable for operation in the range of strong magnetic fields, e.g. those
generated by magnetic resonance imaging!
• The pressure regulator is not fitted with an output pressure gauge. The
connected appliance must be used if you wish to monitor the output pressure
during operation!
Additional regulations apply to some variants of the pressure regulator that supplement
or limit the intended use. The following additional regulations must be followed:
2.2

Additional regulations for pressure regulators with pressure outlet apply

to the series: D44244, D44875, D50440, D55004, D55047
Reduction of 30 MPa (300 bar) inlet pressure maximum to a nominal outlet pressure
of 450 kPa (4.5 bar).
Not suitable for use as a flow control element.
2.3

Additional regulations for pressure regulators with flow outlet apply

to the series: D51735 and D55046
(see section 8)
Reduction of 30 MPa (300 bar) inlet pressure maximum to a maximum outlet
pressure of 500 kPa (5 bar).
For setting the gas flow to the patient's needs.
with flow outlet apply
46
C1/C2

Publicité

Table des Matières
loading

Table des Matières